Abstract

A new, highly sensitive fluorescence immunoassay for a TIRF (total internal reflection)-based point-of-care testing (POCT) device was developed for the detection of procalcitonin (PCT), a specific and early marker for sepsis and microbial infections. The immunoassay was performed on a bench-top system that fulfilled all the necessary characteristics of a POCT application, including a short measurement time (<9min), no sample pre-treatment requirements and application directly near patients. New rat monoclonal antibodies targeting PCT were screened and characterized. The best combinations of antibodies were then integrated into single-use cartridges, and the reduction of nonspecific binding was achieved by supplying suitable additives. Moreover, human recombinant PCT (hrPCT) for use as a standard was developed in the native form of hPCT in plasma (PCT1-116, PCT3-116). The assay achieves the required sensitivity range in human plasma to allow reliable differentiation between healthy persons and varying stages of infection severity (LOD=0.04ng/mL; LOQ=0.12ng/mL). Furthermore, the developed PCT assay can be applied in whole human blood with an adequate sensitivity (LOD=0.02ng/mL; LOQ=0.09ng/mL). To the best of our knowledge, this is the first diagnostic test for sepsis to use whole blood, which is a crucial requirement for POCT. We were able to detect native PCT in patient samples and showed a good correlation (R2=0.988) with the results of the Kryptor® device from BRAHMS, a state of the art device for the detection of PCT.