Abstract

This report represents a 5- to 10-year follow-up of our initial 2- to 5-year data with this proximally hydroxyapatite-coated stem design (Secur-Fit, Stryker Orthopaedics, Mahwah, NJ). This is a retrospective review of a single surgeon's consecutive cases. A total of 105 cases were available for complete review. The average follow-up was 6.7 years (60-123 months). All stems achieved bony ingrowth. Harris hip score and Oxford hip score averaged to 91 and 17, respectively, with no differences seen between the 132° and the 127° stem designs. Leg length measurements were considered equal side-to-side differences of 7 mm or less in 87% of cases: 82% of standard- and 92% of high-offset cases ( P < .05). Four cases of osteolysis were identified and occurred only in patients with an elevated rim liner, implicating neck-liner impingement. The stem design performed exceedingly well over the course of this study. Having a high-offset option aids the surgeon in proper leg length management.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.