Abstract
Total hip arthroplasty in patients requiring very small femoral stems is a challenge because frequently deformity and inadequate bone stock, especially of the acetabulum, are also present. An analysis of 34 consecutive cases managed with a 9-mm distal diameter noncemented stem was performed. The mean patient age was 42.6 years and mean follow-up time was 7.8 years. Two patients were lost to follow-up at 4 and 6 years, and 1 patient had a fracture below the stem immediately after surgery, necessitating revision. These patients were excluded from analysis. Of the remaining 31 patients, 1 underwent revision for distal osteolysis leading to fracture. The Harris hip ratings for these patients were 58.1% excellent, 22.6% good, 16.1% fair, and 3.2% poor. Complications experienced included 1 femoral component that fractured but remains asymptomatic. Polyethylene wear of the very small acetabular components in 31% of cases is a cause for concern; 2 of these patients have required revision for this reason. The acetabular component was subsequently modified to allow thicker polyethylene. Copyright 2003, Elsevier Science (USA). All rights reserved.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.