Abstract

The total electron yield (TEY) mode has been developed successfully for XANES measurements at Beamline 4B7A of BSRF (Beijing Synchrotron Radiation Facility). Its performance was studied by measuring sulphur K-edge XANES of three CdS samples (mixed with graphite powder as an electric conductor) with different concentration: 75%, 50% and 25%. The data are collected in TEY mode and fluorescence yield (FY) mode respectively for comparison. The results demonstrate that the TEY spectra of three samples agree well with each other after the background is subtracted and normalized. The measured XANES spectra by TEY mode without bias and with 100V bias are almost identical to one another, but the signal-to-noise ratio of spectra measured without bias is better than that with 100V bias. The consistency of the self-absorption corrected FY spectra and TEY spectra are within 10% for the three samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.