Abstract
We present a dipping probe total dissolved inorganic carbon (DIC) microsensor based on a localized acidic microenvironment in front of an amperometric CO2 microsensor. The acidic milieu facilitates conversion of bicarbonate and carbonate to CO2, which in turn is reduced at a silver cathode. Interfering oxygen is removed by an acidic CrCl2 oxygen trap. Theoretical simulations of microsensor functioning were performed to find a suitable compromise between response time and near-complete conversion of bicarbonate to CO2. The sensor exhibited a linear response over a wide range of 0-8 mM DIC, with a calculated LOD of 5 μM and a 90% response time of 150 s. The sensor was successfully tested in measuring DIC in bottled mineral water and seawater. This DIC microsensor holds the potential to become an important tool in environmental sensing and beyond for measurements of DIC at high spatial and temporal resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.