Abstract
The torsional properties of a new type of endodontic hand instrument, the Canal Master, were compared with conventional machined K-type endodontic files. Sizes 20 through 50 were tested in clockwise (CW) and counterclockwise (CCW) directions. Canal Master instruments exhibited significantly less torque at yield and at failure in both the CW and CCW directions. Rotation at failure was significantly greater for the Canal Master instruments in both CW and CCW directions. There was no significant difference in either torque or the amount of rotation for the Canal Master instruments when comparing CW versus CCW rotation. There was no clinically visible evidence of deformation of the Canal Master instruments prior to torsional failure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.