Abstract

Titanium (Ti) alloys have been introduced in magnetic resonance (MR) safe implantable medical devices because the susceptibility of Ti is approximately 1/10 that of the Co-Cr-Ni alloy (Elgiloy), which was the previously preferred MR-safe material. The torque applied to metallic materials in an MR imaging (MRI) scanner is commonly believed to increase with the susceptibility of the material. However, a visual inspection showed that the torque applied to Ti alloy cerebral aneurysm clips is comparable with that in the case of those of Elgiloy. In this study, we measured the torque applied to the small test pieces of rods and aneurysm clips quantitatively in a 3-T MRI using an accurate self-developed torque measurement apparatus. The maximum torques of Ti alloy and Elgiloy rod test pieces were comparable as 1.1 and 1.2 µN·m, respectively. The values for Ti alloy aneurysm clips were distinctly higher than the values for those of Elgiloy. These contradictory results of a larger torque for smaller-susceptibility products could be explained by our new theory, which takes into account the crystal susceptibility anisotropy in addition to the conventional torque due to the shape anisotropy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.