Abstract

This paper deals with optimization of a new seal-free compact toroidal coil centrifuge to achieve high resolution in analytical counter-current chromatography (CCC). Toroidal coil CCC (hydrostatic motion) has advantages compared with high-speed CCC (efficiently mixing solution with planetary motion) in the separation of protein or easily emulsified samples. A toroidal coil separation column of 0.4 mm I.D. PTFE tubing was accommodated around the periphery of the cylindrical centrifuge bowl. Using a two-phase solvent system composed of chloroform–acetic acid–0.1 M hydrochloric acid (2:2:1, v/v/v) and a set of dinitrophenyl-amino acids as test samples, a series of experiments was performed with parameters such as the column length, sample volume, flow-rate, elution mode of the mobile phase and rotation speed. The highest efficiency, over 10 000 theoretical plates, was achieved with a 100 m long coiled tube and an 11 ml total capacity at a flow-rate of 0.01 ml/min at 800 rpm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.