Abstract

To report the unexpected induction of astigmatism after phacoemulsification and toric intraocular lens (IOL) implantation in an eye with previous corneal refractive surgery. Case report of a 46-year-old man with bilateral nuclear cataract and previous photorefractive keratectomy. Because corneal topography identified regular corneal astigmatism at the central optical zone, phacoemulsification and implantation of a one-piece hydrophobic acrylic toric IOL were performed. Unexpected induction of astigmatism occurred in the first operated eye despite proper alignment of the IOL according to the preoperative calculations using simulated K values to determine toric IOL power and alignment. A retrospective qualitative analysis of corneal topography showed mismatching of the steepest meridian, leading to an off-axis IOL. Secondary IOL rotation improved both uncorrected and corrected distance visual acuity. Qualitative analysis of the corneal topography is mandatory during the assessment of toric IOL alignment in eyes with previous corneal refractive surgery to identify the actual location of the steepest meridian.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.