Abstract
We report high resolution spectra for the medium molecular weight proteins myoglobin and cytochrome- c obtained using a custom desorption electrospray ionisation (DESI) source coupled to a Bruker Daltonics 12 T Apex Qe Fourier transform ion cyclotron resonance mass spectrometer (FTICR-MS). The DESI source was designed for accurate alignment and reproduction of critical geometric variables. A two axis motorised stage was included to enable automated rastering of the sample under the DESI plume. Spectra for the intact proteins have been obtained under single-acquisition conditions and a top-down analysis of cytochrome- c was performed using both collision induced dissociation (CID) and electron capture dissociation (ECD) of the isolated [M+15H] 15+ charge state. The sequence coverage is comparable to that obtained using electrospray ionisation, demonstrating the utility of top-down protein analysis by DESI FTICR-MS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.