Abstract

A practical method for detecting in vivo cataracts has been developed using a commercially available slit-lamp camera and by utilizing a measuring grid photographic technique. The basis of the method is to photograph a series of slit-beam sections through a lens at varying angles to the optic axis. The areas of opacity show up as localized regions of light back-scatter where the opacities intersect the path of the beam. A set of measuring grids compensated for angle distortion have been prepared. These are superimposed on the appropriate photograph. By taking a sufficient number of sections through a cataractous lens, a tomographic representation of the opacities can be constructed using a lens map. An example of the procedure of three-dimensional mapping is presented using actual lens pictures. The photographic parameters used in this method are explained in detail. The method achieves the goals of ease of use, reproducibility, and applicability to research and clinical studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.