Abstract

We carried on resecrch concerning the characteristics of TLD (Mg_2SiO_4 and MgB_4O_7) in radiotherapy. a) It is necessary to use in measuring radiation dose with 15 MSO_s or 10 MBO_s in order to derive standard error of a range less than ±1.0%. b) There is no dependence of fading within 3 hours and no directional dependence of incident radiation. c) Radiation energy dependence is the most remarkable at 30〜4O keV X-ray and its response exhibits 5-6 times. On other hand, there is no energy dependence at high energy radiation such as a cobalt 60 γ-ray. d) MPO is a better than MSO in exposure dose dependence, and MSO has a poor dose dependence over 100R. e) There is no dependence of depth dose rate at both TLD. As we have seen, if we measure TLD in due consideration of statistics, MBO has better characters than MSO in high energy radiation such as a cobalt 60 γ-ray.