Abstract

The middle ear poses unique challenges when finding suitable materials for ossicular reconstruction, primarily because of its link to the external environment via the eustachian tube, which leads to a greater potential for exposure to infectious agents. In this animal study, the biocompatibilities of titanium and glass-ionomer cement were assessed in the middle ear of the rabbit after being implanted as total ossicular replacement prostheses (TORPs) or as free pins. Animals were sacrified after 28, 84, 168, 336, or 504 days or 2 years, and a cutting saw technique was used to prepare slides for light microscopy. Slides were examined for mucosal coverage and any sign of foreign body reaction. Both materials showed good acceptance in the middle ear. After 28 days, the TORPs were covered by middle ear mucosa. As expected, it took a longer time (up to 504 days) to cover the free implants. An interesting finding was the growth of new bone on both the surface of the titanium implants and the glass-ionomer prostheses. The results of this animal study indicate that both titanium and glass-ionomer cement are favorable materials for ossicular replacement prostheses.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.