Abstract
We present a novel approach to producing bioartificial valves using the tissue-equivalent method of entrapping cells within a biopolymer gel and using a mold design that presents appropriate mechanical constraints to the cell-induced gel compaction to yield both the fibril alignment and the geometry of a native valve. Bileaflet valves were fabricated from bovine collagen and neonatal human dermal fibroblasts as proof of principle. The resultant valves possessed both commissure-tocommissure alignment of collagen fibers in the leaflets and circumferential alignment in the root. While this alignment was manifested in planar biaxial tensile mechanical properties, histology of the leaflets revealed an aligned collagen matrix but lacking other extracellular matrix (ECM) components present in the native valve. The apparent lack of ECM production by the fibroblasts after contracting and aligning the collagen fibrils is consistent with peak loads during biaxial testing being only approximately 10% of native leaflet values and a 0:1 coupling index that was only approximately 50% of native leaflet values despite exhibiting comparable values for the anisotropy index.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.