Abstract

In pigs, latissimus dorsi musculocutaneous island flaps and buttock skin island flaps were raised. Subcutaneous (PscO2) and intramuscular oxygen tension (PimO2) were measured using a non-heated needle electrode before, during and after repeated occlusion of the supplying artery or the draining vein. During arterial and venous occlusion, the tissue oxygen tension in the musculocutaneous flap dropped rapidly. A plateau was reached after 15 min. After arterial occlusion the mean value was 20 mmHg (SEM = +/- 5 mmHg, N = 6) in the subcutis and 16 mmHg in the muscle (SEM = +/- 4 mmHg, N = 10). After venous occlusion the mean value was 11 mmHg (SEM = +/- 3 mmHg, N = 6) in the subcutis. In the skin flap the drop of PscO2 was slower, and after 30 min of arterial occlusion the mean value was 29 mmHg (SEM = +/- 9 mmHg, N = 6). This study has shown that tissue oxygen tension measurement can be used as a sensitive indicator of acute impairment of the supplying vessels in island flaps. The method seems to have potential for monitoring free tissue transfers. A comparable decrease in PscO2 was found for arterial and venous impairment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.