Abstract
This study was undertaken to determine the validity and viability of tissue microarray (TMA) technology in assessing diagnostic immunohistochemistry (IHC) at a single laboratory site. IHC using 57 primary antibodies was performed on a TMA paraffin block containing 89 cores of duplicate previously identified 1 mm diameter tissue specimens. IHC was interpreted by a histology scientist with IHC experience, with pathologist assistance if required. Review of the literature was performed to investigate cases of unexpected immunoreactivity. 55 of 57 antibodies had expected positive staining against the TMA tissue panel that correlated with the original paraffin blocks. Immunostaining of duplicate 1 mm cores correlated with the originally sourced paraffin block in 42 of 43 (98%) tissue types. Some antibodies had unexpected positive immunoreactivity. TMA technology can be utilised effectively in the diagnostic IHC laboratory as a universal positive multiple tissue control block for routine IHC, and can provide valuable information for the benefit of histology scientists and pathologists with respect to interpretation of IHC staining.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.