Abstract
Tissue compensators made with a high atomic number material are compact and retain the skin-sparing effect of megavoltage beams. In this note, we have described the use of a commercially available vinyl lead as a material for tissue compensators for use with 4-MV beams on a Clinac-4/80. The dosimetric data obtained included the sheet's lead equivalence, thickness ratios, surface dose and buildup region, and transmission factors. The presence of a compensator did not alter the beam's skin-sparing effects. It was concluded that the vinyl lead sheets allowed an easy and rapid fabrication of a tissue compensator for head and neck portals of up to 15 X 15 cm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.