Abstract

In this paper, the impact of the tip radius on dopant profiling by scanning microwave microscopy is investigated. The cantilevers are very likely to erode in such measurements, and thus, a two-dimensional Poisson solver was used to calculate the lateral spatial resolution as a function of tip radius and doping. Moreover, a strong correlation between the slope of the calibration curves and the tip diameter was found. The slope of the calibration curves increases toward −0.5 and saturates as the radius approaches values of 150 nm, which is in agreement with experimental data obtained from scanning microwave microscopy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.