Abstract

BACKGROUNDBloodstream infections are associated with high rates of morbidity and mortality. Rapid detection of bloodstream infections is important in achieving better patient outcomes.OBJECTIVECompare the time-to-detection (TTD) of the new BacT/Alert Virtuo and the BACTEC FX automated blood culture systems.DESIGNProspective simulated comparison of two instruments using seeded samples.SETTINGMedical microbiology laboratory.METHODSBlood culture bottles were seeded in triplicate with each of the standard ATCC strains of aerobes, anaerobes and yeast. TTD was calculated as the length of time from the beginning of culture incubation to the detection of bacterial growth.MAIN OUTCOME MEASURESTTD for the various tested organisms on the two microbial detection systems.RESULTSThe 99 bottles of seeded blood cultures incubated in each of the blood culture systems included 21 anaerobic, 39 aerobic and 39 pediatric bottles. The BacT/Alert Virtuo system exhibited significantly shorter TTD for 72.7 % of the tested organisms compared to BACTEC FX system with a median difference in mean TTD of 2.1 hours (interquartile range: 1.5–3.5 hours). The BACTEC FX system was faster in 15.2% (5/33) of microorganisms, with a median difference in mean TTD of 25.9 hours (IQR: 9.1–29.2 hours).CONCLUSIONTTD was significantly shorter for most of the microorganisms tested on the new BacT/Alert Virtuo system compared to the BACTEC FX system.LIMITATIONSUse of simulated cultures to assess TTD may not precisely represent clinical blood cultures.