Abstract

In this study, a novel measurement method for a secondary ultrasound wave irradiated by microbubble cavitation is proposed. High-intensity ultrasound (h-US, 1.0–1.5 MPa), which produces bubble cavitation, is irradiated with a fixed time delay after introducing imaging US, whose frequency is different from that of the h-US. The bubble cavitation signal (BCS) is detected by the signal-processing unit of an ultrasound power Doppler imaging instrument. By this method, both a spatially resolved bubble image (S-image) and the temporal transition of the BCS (T-image) are monitored simultaneously. A feature of the method is that the BCS is observed in situ with sub-µs time resolution. The accuracy of the method is evaluated and it is found that the maximum deviation of the amplitude of the simulated BCS is 4.80%. This method is applied to measure the BCS of ultrasound contrast agent microbubbles. As a result, the dependence of the inherent temporal transition of the BCS on the sound pressure of the h-US (0.6–1.2 MPa) is observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.