Abstract

Intracavity laser absorption spectroscopy (ICLAS) with an evacuated Cr2+:ZnSe laser is performed with a high-resolution time-resolved Fourier transform interferometer with a minimum detectable absorption coefficient equal to 4×10−9 cm−1Hz−1∕2 in the 2.5 μm region. This represents the extreme limit currently reached in the infrared by ICLAS with Doppler-limited resolution. The broad gain band of the crystal allows a spectral coverage at most equal to 125 nm, wide enough to see entire vibration bands. Weak CO2 bands observed up to now only in the Venusian atmosphere are recorded for the first time, to our knowledge, in a laboratory. An H2O detection limit down to 0.9 parts per billion by volume is also demonstrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.