Abstract

A depth profile analysis of SiO2 glass ion implanted with 23Na+ was carried out on a time‐of‐flight secondary ion mass spectrometry (ToF‐SIMS) with OX, Cs and buckminsterfullerene (C60) ion sputtering. The Na migration in SiO2 glass during both analysis process and sputter process was investigated under the various operating conditions. During the analysis process, 23Na+ intensity stayed constant on the dose density of the Bi primary ion regardless of the conditions and went up using higher primary ion current. The Na profile in SiO2 glass similar to that of the Lindard–Scharff–Schiott theory was successfully acquired by the use of C60 ion sputtering regardless of the operating conditions. It is considered that Na does not migrate during the sputter process with C60 sputter ion beam. In addition, the detection limit of Na depth profile with C60 ion sputtering was almost the same with existing dynamic SIMS analysis with Cs primary ion. Thus, the ToF‐SIMS with C60 ion sputtering is shown to be more versatile than the existing method by avoiding the necessity to optimize the operating conditions. Copyright © 2012 John Wiley & Sons, Ltd.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.