Abstract

Based on the adsorption of analytes in the sampling loop, a time-resolved dynamic dilution introduction method was developed for negative ion mobility spectrometry to continuously monitor end-tidal propofol without other sample pre-separation. The dynamic dilution characteristics of propofol and moisture in the Teflon sample loop (4 mm o.d. and 2.4 mm i.d., 150 cm length) were both theoretically and experimentally investigated. The prominent absorption differences between propofol and moisture on the inwall of the sample loop allowed their concentrations to be time-resolved during the injection process, realizing sensitive measurement of end-tidal propofol with a response time of 2 s. At the optimized carrier gas flow rate of 700 mL min−1, the linear response range for propofol was achieved to be 0.2 to 20 ppbv with a limit of detection (LOD) of 65 pptv. Finally, this method was performed on a patient undergoing mastectomy surgery to continuously monitor the end-tidal propofol with an interval of five respirations and the result nicely demonstrated its fast response to the propofol changes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.