Abstract
We present time-resolved radioluminescence (RL) measurements from P-doped silica optical fibre, demonstrating potential utility in pulsed source dosimetry. When subjected to 140 MU/min from a 6 MV photon linac source, a 220 µm-core fibre has produced a RL response of 720 ± 20 photon counts/pulse from a saw-tooth ~ 40 µs duration return-to-baseline waveform. Conversely, the Cerenkov stem signal within the radiation insensitive carrier fibre is observed to offer an amplitude amounting to a little less than 3% of that of the P-doped fibre, the sharply spiked response being of ~ 2 µs duration. On the basis of these results, the practical applications and the challenges in the establishment of an effective dosimetry system are discussed.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.