Abstract
We demonstrate the feasibility of time-resolved diffuse reflectance measurements at small source-detector separations using a single-photon avalanche diode operated in time-gated mode. Photon time distributions at an interfiber distance of 2 mm were obtained on a homogeneous tissue phantom with a dynamic range of 10(6) and collecting photons at arrival times up to 4 ns. Moreover, we were able to detect a local inhomogeneity deeply buried within a diffusive medium with better spatial resolution, higher signal intensity, and same contrast of a larger (20 mm) interfiber distance. Finally, the proposed approach proved valuable to detect in vivo a task-related brain activation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.