Abstract

The application of digital sampling technique for extracting the timing properties of parallel-plate avalanche counter (PPAC) is described. The preamplifier output signals of two identical avalanche counters were simultaneously digitized by using a digital storage oscilloscope (sampling rate up to 10 GS/s, 8-bit resolution) and analyzed offline. The offline analysis includes signal and noise discrimination, noise filtration and determination of pulse arrival time. A time resolution better than 398 ps was obtained which matches the best literature results using analog electronics. The effect of sampling rate of signals on the achievable time resolution is studied as well. The results of the investigation lead to a simple and flexible method of data analysis in experiments dealing with avalanche counters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.