Abstract

Objective.Time-of-flight (TOF) scatter rejection allows for identifying and discarding scattered photons without the use of an anti-scatter grid (ASG). Although TOF scatter rejection was initially presented for cone-beam computed tomography, we propose, herein, to extend this approach to x-ray radiography. This work aims to evaluate with simulations if TOF scatter rejection can outperform ASGs for radiography.Approach.GATE was used to simulate the radiography of a head and a torso and a water cylinder with bone inserts in a system with total timing jitters from 0 ps up to 500 ps full-width-at-half-maximum. The transmission factor of TOF scatter rejection for primary and scattered photons was evaluated as if it were a virtual ASG.Main results.With a total timing jitter of 50 ps, TOF scatter rejection can reach a selectivity of 4.93 with a primary photons transmission of 99%. Reducing the timing jitter close to 0 ps increases the selectivity up to 15.85 for a head and torso radiography, outperforming typical ASGs which usually have a selectivity from 2.5 to 10 with a primary photons transmission from 50% to 70%.Significance.This suggests that TOF scatter rejection may be suitable to replace ASGs in applications requiring lower radiation exposure if sufficiently low timing jitter is achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.