Abstract

An application of digital sampling techniques is presented which can simplify experiments involving sub-nanosecond time-mark determinations and energy measurements with nuclear detectors, used for Pulse Shape Analysis and Time of Flight measurements in heavy ion experiments. The basic principles of the method are discussed as well as the main parameters that influence the accuracy of the measurements. The method allows to obtain both time and amplitude information with an electronic chain simply consisting of a charge preamplifier and a fast high resolution ADC (in the present application: 100 MSample/s , 12 bit ) coupled to an efficient on-line software. In particular an accurate Time of Flight information can be obtained by mixing a beam related time signal with the output of the preamplifier. Examples of this technique applied to Silicon detectors in heavy-ions experiments involving particle identification via Pulse Shape analysis and Time of Flight measurements are presented. The system is suited for applications to large detector arrays and to different kinds of detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.