Abstract

This paper presents a method that corrects pulse waveforms distorted by the frequency-dependent loss of microwave cables in measuring pulsed electric fields (PEFs). Because the distortion resulting from the microwave cable disrupts accurate PEF measurements, the distorted pulse should be corrected for precise PEF effect testing. The proposed correction method is achieved by a transfer function that is determined by ABCD parameters calculated from the scattering parameters of the cable. A 10-m microwave cable is tested to validate the proposed method, where the input pulse is a 2-ns sine pulse of a single cycle. Here, the output pulse, scattering parameters, and cable resistance are measured. These measurement results are used to represent the transfer function in MATLAB for the proposed correction method. The test results show that the corrected pulse obtained from the transfer function has an error of 4.5% in the peak-to-peak voltage and an error of 0.8% in the bipolar pulse width compared to the reference input pulse. The errors of PEF measurement decrease dramatically by using the proposed correction method. Moreover, the correction method is validated for various pulse durations, pulse shapes, and cable types.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call