Abstract

High-temperature, high-density experiments require a simultaneous understanding of temporal and spectral regions. The spectral x-ray streak camera (SXSC) is a new high-temporal-resolution spectral x-ray diagnostic system that allows researchers to differentiate between soft and hard x-ray regions. The diagnostic offers three spectral channels with a wide spectral range, one direct channel that includes a filter and two indirect channels that include both mirrors and filters. The opto-mechanical design positions the filtered radiation at three different locations along the streak photo-cathode (PC) slit to provide time-dependent spectral channels with pico-second temporal resolution. A moderate spatial resolution (150-700 μm) is achieved using slits perpendicular to the PC slit, while the slit width is optimized according to the central channel wavelength (for each channel). The diagnostic system covers a spectral range of 30-500eV for the mirror channels and >1300eV for the direct channel. The temporal and spatial axes of the streak camera are calibrated with respect to a sequence of x-ray pulses. The SXSC diagnostic system is tested and analyzed using Marshak-wave emission from an SiO2 foam that was heated by a laser-beam irradiated halfraum. The SXSC results are compared to measurements from an x-ray diode array with similar spectral channels.