Abstract

BackgroundVentilator-associated pneumonia (VAP) is the most common nosocomial infection in intensive care units. Distal airway mucus clearance has been shown to reduce VAP incidence. Studies suggest that mucus clearance is enhanced when the rate of expiratory flow is greater than inspiratory flow. The time-controlled adaptive ventilation (TCAV) protocol using the airway pressure release ventilation (APRV) mode has a significantly increased expiratory relative to inspiratory flow rate, as compared with the Acute Respiratory Distress Syndrome Network (ARDSnet) protocol using the conventional ventilation mode of volume assist control (VAC). We hypothesized the TCAV protocol would be superior to the ARDSnet protocol at clearing mucus by a mechanism of net flow in the expiratory direction.MethodsPreserved pig lungs fitted with an endotracheal tube (ETT) were used as a model to study the effect of multiple combinations of peak inspiratory (IPF) and peak expiratory flow rate (EPF) on simulated mucus movement within the ETT. Mechanical ventilation was randomized into 6 groups (n = 10 runs/group): group 1—TCAV protocol settings with an end-expiratory pressure (PLow) of 0 cmH2O and PHigh 25 cmH2O, group 2—modified TCAV protocol with increased PLow 5 cmH2O and PHigh 25 cmH2O, group 3—modified TCAV with PLow 10 cmH2O and PHigh 25 cmH2O, group 4—ARDSnet protocol using low tidal volume (LTV) and PEEP 0 cmH2O, group 5—ARDSnet protocol using LTV and PEEP 10 cmH2O, and group 6—ARDSnet protocol using LTV and PEEP 20 cmH2O. PEEP of ARDSnet is analogous to PLow of TCAV. Proximal (towards the ventilator) mucus movement distance was recorded after 1 min of ventilation in each group.ResultsThe TCAV protocol groups 1, 2, and 3 generated significantly greater peak expiratory flow (EPF 51.3 L/min, 46.8 L/min, 36.8 L/min, respectively) as compared to the ARDSnet protocol groups 4, 5, and 6 (32.9 L/min, 23.5 L/min, and 23.2 L/min, respectively) (p < 0.001). The TCAV groups also demonstrated the greatest proximal mucus movement (7.95 cm/min, 5.8 cm/min, 1.9 cm/min) (p < 0.01). All ARDSnet protocol groups (4–6) had zero proximal mucus movement (0 cm/min).ConclusionsThe TCAV protocol groups promoted the greatest proximal movement of simulated mucus as compared to the ARDSnet protocol groups in this excised lung model. The TCAV protocol settings resulted in the highest EPF and the greatest proximal movement of mucus. Increasing PLow reduced proximal mucus movement. We speculate that proximal mucus movement is driven by EPF when EPF is greater than IPF, creating a net force in the proximal direction.

Highlights

  • Ventilator-associated pneumonia (VAP) is the most common nosocomial infection in intensive care units

  • The time-controlled adaptive ventilation (TCAV) protocol settings resulted in the highest expiratory flow rate (EPF) and the greatest proximal movement of mucus

  • We speculate that proximal mucus movement is driven by EPF when EPF is greater than Peak inspiratory flow (IPF), creating a net force in the proximal direction

Read more

Summary

Introduction

Ventilator-associated pneumonia (VAP) is the most common nosocomial infection in intensive care units. Studies suggest that mucus clearance is enhanced when the rate of expiratory flow is greater than inspiratory flow. Retention of respiratory mucus in the distal airways has been implicated as a pathogenic mechanism of VAP; it impairs immunological responses, alters normal respiratory physiology, and is highly associated with the development of pneumonia [3,4,5,6]. Mucus clearance is enhanced with ventilator settings in which the expiratory (outward) flow rate is greater than the inspiratory flow rate.

Objectives
Methods
Results
Discussion
Conclusion