Abstract
The detection efficiency η of any particle detector is important, concerning acquisition time, but becomes even more critical when two particles are detected in coincidence, with a total efficiency η1η2, in order to allow a deeper understanding of complex processes induced by light or particle interaction with matter. Efficiency and resolution of a time and position sensitive x-ray detector are reported here. This system consists of a multilayer transmission photocathode and two micro-channel plates (MCPs) equipped with a delay line anode (DLA). The efficiency is found to be about 20% for Al Kα photons, while the spatial resolution is comparable to that of a standard DLA detector (about 100 μm). The fast response time of the detector combined with its efficiency should allow coincidence experiments between x-ray photons and other particles (electron, ions, etc.).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.