Abstract

Contrast transfer function (CTF) estimation is a necessary step in the cryo-electron tomography (cryoET) workflow and essential for high-resolution in situ structural determination. However, the low signal-to-noise ratio and continuous defocus variation in micrographs of cryoET tilt series make accurate CTF estimation challenging. Here, we report a tilt-series-based joint CTF estimation method implemented in the new software CTFMeasure. The joint estimation method combines all Thon-ring signals in a tilt series to improve the estimation accuracy. By using an objective function involving the CTF parameters and geometric parameters of a cryoET tilt series, CTFMeasure can estimate the CTF parameters of each micrograph and the absolute tilt angle offset of the lamellar sample relative to the sample stage plane, which is usually the glancing angle used during focused ion beam (FIB) milling. Tests on both synthetic and experimental data, as well as subtomogram averaging, demonstrated the accurate CTF estimation of cryoET tilt series by CTFMeasure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.