Abstract

The image quality in cone-beam computed tomography (CBCT) should be improved tilting the mandible that contains two dental titanium implants, within the relevant range of motion. Using the mandible of a five-month-old pig, CBCT was performed varying the accelerating voltage, beam current, the starting rotation angle of the mandible in the source-detector plane and the tilt angles of the jaw with respect to the source-detector plane. The different datasets were automatically registered with respect to micro CT data to extract the common volume and the deviance to the pre-defined standard that characterizes the image quality. The variations of the accelerating voltage, beam current and the rotation within the source-detection plane provided the expected quantitative behavior indicating the appropriate choice of the imaging quality factor. The tilting of the porcine mandible by about 14° improves the image quality by almost a factor of two. The tilting of the mandible with two dental implants can be used to significantly reduce the artifacts of the strongly X-ray absorbing materials in the CBCT images. The comparison of 14° jaw tilting with respect to the currently recommended arrangement in plane with the teeth demonstrates that the applied exposure time and the related dose can be reduced by a factor of four without decreasing the image quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.