Abstract

Background Airway obstruction can occur during facemask (FM) resuscitation of preterm infants at birth. Intubation bypasses any upper airway obstruction. Thus, it would be expected that the occurrence of low expiratory tidal volumes (VTes) would be less in infants resuscitated via an endotracheal tube (ETT) rather than via an FM. Our aim was to test this hypothesis. Methods Analysis was undertaken of respiratory function monitoring traces made during initial resuscitation in the delivery suite to determine the peak inflating pressure (PIP), positive end expiratory pressure (PEEP), the VTe and maximum exhaled carbon dioxide (ETCO2) levels and the number of inflations with a low VTe (less than 2.2 mL/kg). Results Eighteen infants were resuscitated via an ETT and 11 via an FM, all born at less than 29 weeks of gestation. Similar inflation pressures were used in both groups (17.2 vs. 18.8 cmH2O, P = 0.67). The proportion of infants with a low median VTe (P = 0.6) and the proportion of inflations with a low VTe were similar in the groups (P = 0.10), as was the lung compliance (P = 0.67). Infants with the lowest VTe had the stiffest lungs (P < 0.001). Conclusion Respiratory function monitoring during initial resuscitation can objectively identify infants who may require escalation of inflation pressures.