BackgroundFor adequate ventilation during bag-valve-mask ventilation, rescuers should ensure a proper mask seal using the one-handed or two-handed technique. Little is known about the magnitude of sealing forces of a bag-valve-mask needed for adequate ventilation. This study aimed to explore the effect of the 4-point sealing forces of a bag-valve-mask on tidal volume while using the one-handed technique, focusing on the moderating effect of C length (the distance from the thumb to the index finger in the C shape of the one-hand EC grip).MethodsA prospective, descriptive simulation study was conducted. A convenience sample of 125 undergraduate paramedic students from two universities was recruited. A self-reported questionnaire was used to collect subjective variables. Tidal volumes, 4-point sealing forces of the mask, peak pressure, and C length of the C shape in the one-hand EC grip were measured using the mechanical lung model under a simulated adult respiratory arrest. Hierarchical regression analysis was used to determine the moderating effect of C length on tidal volume in bag-valve-mask ventilation.ResultsThe average C length, peak pressure, and tidal volume were 7.54 (± 1.85) cm, 11.62 (± 5.40) cmH2O, and 321.66 (± 135.18) mL, respectively. The average range of the 4-point sealing forces was 0.03–0.69 N. The apex sealing force was the weakest among the 4-point sealing forces. Hierarchical regression analysis demonstrated that tidal volume accounted for 62.7% of the variance by C length, peak pressure, and apex sealing force during bag-valve-mask ventilation (F = 9.676, p < 0.001). C length moderated the effect of the apex sealing force and peak pressure on the tidal volume, meaning the higher the peak pressure and apex sealing force, the more the tidal volume and the longer the C length.ConclusionThis first simulation study measuring the 4-point sealing forces during bag-valve-mask ventilation provides effective advice that can be adopted in clinical practice without side effects and underpins the importance of continuous retraining and assessment focused on individual physical characteristics, such as C length and bag-valve-mask sealing forces.

Highlights

  • For adequate ventilation during bag-valve-mask ventilation, rescuers should ensure a proper mask seal using the one-handed or two-handed technique

  • Rescuers seal the mask against the face using either one- or two-handed techniques, with the thumb and the index finger wrapped in a “C” shape around the mask apex, and the remaining fingers in an “E” shape, lifting the jaw [5]; this is called the “EC grip”

  • tidal volume (TV) had a negative correlation with C length (r = − 0.223, p = 0.012)

Read more

Summary

Introduction

For adequate ventilation during bag-valve-mask ventilation, rescuers should ensure a proper mask seal using the one-handed or two-handed technique. This study aimed to explore the effect of the 4-point sealing forces of a bag-valve-mask on tidal volume while using the one-handed technique, focusing on the moderating effect of C length (the distance from the thumb to the index finger in the C shape of the one-hand EC grip). Adequate ventilation involves delivering each rescue breathing over 1 s to provide a sufficient tidal volume (TV), using the BVM, and produce a visible chest rise [1]. This is deeply associated with the survival rate as well as oxygenation of patients with cardiac or respiratory arrest [2]. Several previous studies [7,8,9,10,11,12,13,14,15,16,17] have reported that the TV delivered during BVM ventilation is affected by rescuers’ squeezing methods, such as the thenar eminence technique, E-O technique, and E-C technique, as well as individual characteristics, such as gender, height, weight, hand width and length, grip power, education, practice, and experience, among others; the reported results have been inconsistent

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.