Abstract

Ti-C:H were deposited on untreated Ti6Al4V alloy and high temperature nitride Ti6Al4V alloy by Closed Field Unbalanced Magnetron Sputtering (CFUMBS). Therefore, four type of specimens, Ti6Al4V, nitride Ti6Al4V(N- Ti6Al4V), T-C:H deposition on former specimens (T-C:H/Ti6Al4V, T-C:H/ N-Ti6Al4V). The characterization of the Ti-C:H film behavior was studied by means of XRD, SEM, scratch tester, nanoindenter. To simulate human body environment, wear experiments were performed by using Schwingung Reibungund Verschliss(SRV) wear tester in 0.64% NaCl solution. Corrosion performance of Ti-C:H coatings was investigated by potentio statin strument (Jiehan ECW-5000,Taiwan). All tests were carried out in a 0.64% NaCl solution. To evaluate the biocompatibility of the specimens were cultured by Purified mouse leukaemic monocyte macrophage cells (Raw 264.7) and variations of cell morphology following various periods were observed. The result indicate that Ti-C:H film show great wear residence. In addition , Ti-C:H film have good corrosion resistance and biocompatibility in this study.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.