Abstract

ObjectivesWe assessed the analytical performance of the TSH and FT4 assays on ADVIA Centaur in a multicenter national evaluation. Design and methodsA precision study and a method comparison were performed. Reference values stated by the manufacturer were checked from 379 normal subjects. ResultsFor TSH and FT4, the intra-assay CVs were below 2.3 and 5.2%, respectively, and the inter-assay CVs below 4.4% and 7.2%, respectively. Therefore, the precision and reproducibility were acceptable. Bland–Altman bias plots revealed good correlation and agreement with Cobas assays. TSH and FT4 data yielded reference ranges of 0.64–3.24mIU/L and 10.5–18.9pmol/L, respectively. ConclusionThese assays demonstrate reliable characteristics. The reference ranges obtained can be used for interpretation of thyroid function.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.