Abstract

The sites of thrombus formation in the Jarvik III artificial heart were studied in 20 consecutive calf experiments. Identical design and implantation procedures were used with two different surfaces. The 10 anticoagulated calves receiving Dacron-fibrilized silicone rubber surfaces survived an average of 296 hours. In contrast, the 10 nonanticoagulated calves with smooth poyurethane hearts survived 545 hours. The polyurethane surfaces were constantly clean. However, the rough surfaces were coated with variable amounts of thrombi. The sharp angle between the diaphragm and housing contained varying amounts of thrombi deposits in 75 per cent of the cases. Thrombe deposition occurred on 41 per cent of the valves. These thrombe were the source of emboli into the kidney, brain, and lungs. Turbulences and stagnation areas generated thrombus formation. Anticoagulation did not prevent thrombus formation. It would appear that heart free of turbulence and stagnation areas must be designed to eliminate thromboembolism.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.