Abstract

Postimplant device thrombosis remains a life-threatening complication and limitation of continuous-flow ventricular assist devices (VADs). Using advanced computational fluid dynamic (CFD) simulations, we successfully depicted various flow patterns, recirculation zones, and stagnant platelet trajectories which promote thrombus formation and observed that they matched actual thrombus formation patterns observed in Thoratec HeartMate II VADs explanted from patients with pump thrombosis. Previously, these small eddies could not be captured by either digital particle image velocimetry or CFD due to insufficient resolution. Our study successfully demonstrated the potential capability of advanced CFD to be adopted for device optimization, leading to enhanced safety and efficacy of VADs for long-term destination therapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.