Abstract

Three-dimensional (3-D) CT volumetric filling ratio (VFR) is a better predictor of proximal humerus stress shielding after short-stem anatomic total shoulder arthroplasty (aTSA) than using plain radiographs. Forty-four patients with short-stem aTSAs, preoperative CT scans, and a minimum 3-year radiographic follow-up were included. Patients were divided into group A (stress shielding) and group B (no stress shielding) based on the radiographic analysis. Standard implant filling ratios were measured on plain radiographs. The 3-D VFR of the metaphyseal and diaphyseal segments of the aTSA stem was measured using MIMICS (Materialise). The area under a receiver operator characteristic curve was used to determine the predictive strength of the 3-D VFR method. The average age and radiographic follow-up was 69 years and 44 months. Group A had 19 patients and statistically higher filling ratios using 3-D VFR method than group B, whereas no notable differences were found between the groups using standard techniques. The 3-D VFR had an area under a receiver operator characteristic curve of 92%, which supports it as a good predictor of stress shielding. These methods enable early identification of patients at risk for stress shielding and can also be valuable in improving humeral stem designs. Level III. Case-control study.