Abstract
Clinical use of image-guided needle placement robots has lagged behind laboratory-demonstrated robotic capability. Bridging this gap requires reliable and easy-to-use robotic systems. Our system for image-guided needle placement requires only simple, low-cost components and minimal, entirely off-line calibration. It rapidly aligns needles to planned entry paths using 3D ultrasound (US) reconstructed from freehand 2D scans. We compare system accuracy against clinical standard manual needle placement. The US-guided robotic system is significantly more accurate than single manual insertions. When several manual withdrawals and reinsertions are allowed, accuracy becomes equivalent. In ex vivo experiments, robotic repeatability was 1.56 mm, compared to 3.19 and 4.63 mm for two sets of manual insertions. In an in vivo experiment with heartbeat and respiratory effects, robotic system accuracy was 5.5 mm. A 3D US-guided robot can eliminate error bias and reduce invasiveness (the number of insertions required) compared to manual needle insertion. Remaining future challenges include target motion compensation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The International Journal of Medical Robotics and Computer Assisted Surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.