Abstract

Two-dimensional (2D) Look-Locker MRI technique can identify myocardial fibrosis, but cannot cover the whole left ventricle during a single scan. To develop breath-hold three-dimensional (3D) Look-Locker MRI for the evaluation of postcontrast myocardial and blood T1 values and myocardial scarring in the left ventricle. A phantom and 24 patients with myocardial diseases underwent gadolinium-enhanced 2D and 3D Look-Locker MRI using a 1.5-T unit. We compared the T1 value of the phantom and the values of the myocardium and blood in the patients between the two Look-Locker MRI sequences. In the patient study, the scan ordering of the two Look-Locker MRI was selected randomly. We also assessed the ability of the 3D imaging to detect myocardial scarring that was confirmed by late gadolinium enhancement MRI. The phantom study showed a good agreement for the T1 value between 2D and 3D Look-Locker MRI. There were no significant differences in the myocardial T1 values after contrast between 2D and 3D Look-Locker MRI or in the T1 values between the two imaging slices on the 3D Look-Locker MRI (P > 0.10). A better agreement for the myocardial T1 values was found when the 3D Look-Locker imaging was performed first. The T1 values for blood were affected by the scan order (P < 0.05). The 3D Look-Locker MRI showed myocardial scarring with a shorter T1 value (290.4 ± 62.7 ms) than those for unscarred myocardium (360.8 ± 30.3 ms). Three-dimensional Look-Locker MRI may precisely estimate the postcontrast myocardial and blood T1 values for the entire left ventricle during a single scan.