Abstract
A three-dimensional iterative reconstruction algorithm which incorporates models of the geometric point response in the projector-backprojector is presented for parallel, fan, and cone beam geometries. The algorithms have been tested on an IBM 3090-600S supercomputer. The iterative EM reconstruction algorithm is 50 times longer with geometric response and photon attenuation models than without modeling these physical effects. An improvement in image quality in the reconstruction of projection data collected from a single-photon-emission computed tomography (SPECT) imaging system has been observed. Significant improvements in image quality are obtained when the geometric point response and attenuation are appropriately compensated. It is observed that resolution is significantly improved with attenuation correction alone. Using phantom experiments, it is observed that the modeling of the spatial system response imposes a smoothing without loss of resolution. >
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.