Abstract

Discrepancy between maximum diameters obtained with two-dimensional ultrasound and computed tomography (CT) after endovascular aneurysm repair (EVAR) is well known. The maximal diameter is ideally measured perpendicular to the centerline, a methodology so far only feasible with three-dimensional (3D) CT and magnetic resonance angiography (MRA). We aimed to investigate the agreement between 3D ultrasound and 3D CT and to determine reproducibility measures. Prospective study comparing 3D ultrasound with 3D CT in 124 consecutive patients seen 3 or 12 month after EVAR. Replacing 2D with 3D ultrasound, the mean difference was improved from 6.0 mm to -1.3 mm (p < .001), and the range of variability was reduced from 9.4 mm to 6.6 mm (p = .009) using 3D CT as the gold standard. The mean difference between 3D ultrasound and 3D CT maximum diameter of the residual sac was -1.3 mm with upper and lower limits of agreement of 5.2 mm and -7.9 mm, respectively. Reproducibility measures of 3D ultrasound were ± 4 mm. 3D ultrasound correlate significantly better to 3D CT than the currently used 2D ultrasound method when assessing maximum diameter of the residual sac after EVAR, and reproducibility measures were within clinical acceptable values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.