Simple experiments which form the basis for a true 3-D demonstration of sectional images are presented and a method for genuine 3-D display of dynamic ultrasound images is described. Eight ultrasound images are recorded with a slightly different angulation of the transducer. The images are extracted from the video signal from a conventional ultrasound scanner and stored in eight digital memories. After recording, each image is displayed on an oscilloscope screen, which is viewed via a fast oscillating mirror. The position of the mirror determines which of the eight images are to be displayed and thereby ensures a correct spatial relationship of the images, resulting in a true 3-D scan presentation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.