Abstract
We applied photoacoustics as a tissue tomography technique for the detection of blood concentrations, e.g., angiogenesis around tumors. We imaged blood vessels in highly scattering samples, using 532-nm light, to depths of ~1 cm . The samples were real tissue (chicken breast) or 10% dilutions of Intralipid-10%. The blood flowed through nylon capillaries. Polyvinylidene difluoride (PVdF) piezoelectric detectors were used in a surface-scanning mode. We demonstrate the sensitivity of the technique by photoacoustic detection of single red blood cells upon a glass plate. Lateral resolution is limited by the detector diameter (200 microm). The depth resolution is ~10 microm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.