Abstract
The aim of this work was to investigate the suitability of a novel forward-viewing intravascular ultrasound (IVUS) technique for three-dimensional imaging of severely stenosed or totally occluded vessels, where the conventional side-viewing IVUS systems are of limited use. A stiff 3.8 mm diameter forward-viewing catheter was manufactured to scan a 72° sector ahead of its tip. Conical volume data were acquired by rotating the catheter over 180° by means of a motorised mechanical system. Operating at 30 MHz, the catheter was integrated with an IVUS scanner and a radiofrequency data acquisition system. Postmortem carotid and femoral arteries were scanned in vitro. Correlation of the reconstructed images with histology demonstrated the ability of this forward-viewing IVUS system to visualise healthy lumens, bifurcations, thickened atherosclerotic walls and, most importantly, severe and complete vessel occlusions. A rotating-sector forward-viewing IVUS system is suitable for anatomical assessment of severely diseased vessels in three dimensions. (E-mail: N.McDicken@ed.ac.uk)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.