Abstract

The first true three-dimensional image reconstructions from a metal-walled vessel using electrical impedance tomography (EIT) are presented. Two image reconstruction techniques have been applied via relatively sophisticated FEM modelling of a bespoke laboratory test vessel from which data have been obtained using an EIT instrument designed to intrinsically safe requirements. A generalized Tikhonov regularization method is compared with the linear back-projection (LBP) technique. Subsequent image reconstructions strongly suggest that the LBP method when applied to a metal-walled vessel is highly sensitive to the level of detail within the FEM model. By comparison, the regularized technique is far less sensitive to the complexity of the modelled geometry. Additionally, unlike the LBP method, the regularization technique has been successful in accurately reconstructing multiple inhomogeneities within an aqueous system. A further experiment has shown similar sensitivity in a wetted powder-based system. It is concluded that EIT via a regularized difference imaging approach has significant potential for detecting 3D malformations and non-uniformities in industrial pressure filtration systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.