Abstract
Thiol-functionalized reduced graphene oxide (TRGO) as a novel ion-to-electron transducing layer is firstly employed to develop durable solid-contact ion-selective electrodes (SC-ISEs) in this work. The performance of the sensors is evaluated by determining K+ and NO3− as an example of cation and anion. The covalent linkage of TRGO at golden electrode surface generates a stable transducing layer. No water films are observed in the proposed TRGO-based potassium (K+-TRGO-ISEs) and nitrate (NO3−-TRGO-ISEs) selective SC-ISEs. The resultant electrodes exhibit Nernstian responses (60.0 ± 0.4 mV/decade for K+-TRGO-ISEs and −60.0 ± 0.5 mV/decade for NO3−-TRGO-ISEs), low detection limits (2.5 × 10−6 M for K+-TRGO-ISEs and 4.0 × 10−6 M for NO3−-TRGO-ISEs) and good selectivity behavior. More importantly, the TRGO-based SC-ISEs display a much longer lifetime of 2 weeks than that of reduced graphene oxide-based SC-ISEs in continuous flowing solutions using a longer peristaltic pump. These improvements push TRGO a general and reliable transducer for the development of durable SC-ISEs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Talanta
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.